Back to Search Start Over

Feasibility Study of a Novel Navigation System for Biopsy of Peripheral Lesions in the Lungs.

Authors :
GRUIONU, L. G.
SĂFTOIU, A.
POPA, T.
CIOBÂRCĂ, C.
STREBA, C. T.
IONCICĂ, ANA-MARIA
GRUIONU, G.
Source :
Current Health Sciences Journal; Jan-Mar2016, Vol. 42 Issue 1, p76-81, 6p
Publication Year :
2016

Abstract

A novel system for electromagnetic navigation in bronchoscopy (ENB) to improve peripheral lesion targeting and diagnostic is currently under development, and preliminary tests on a complex phantom are presented in this paper. The system named Transbite is proving to be easy to use, accurate and useful for experienced and beginner users, with precision around 5-9 mm in reaching targets in sub-segmental bronchi where a conventional video-bronchoscope cannot extent. Transbite system consists of a planning and orientation software, a navigation and biopsy forceps, an active marker that is placed on patient chest and an electromagnetic tracking system from the market connected to a computer containing the software. Transbite can be used with any bronchoscopy system, has a short set-up procedure and learning curve. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
20670656
Volume :
42
Issue :
1
Database :
Complementary Index
Journal :
Current Health Sciences Journal
Publication Type :
Academic Journal
Accession number :
118351688
Full Text :
https://doi.org/10.12865/CHSJ.42.01.11