Back to Search Start Over

A Coil Constant Calibration Method Based on the Phase-Frequency Response of Alkali Atomic Magnetometer.

Authors :
Yao, Han
Ma, Danyue
Zhao, Junpeng
Lu, Jixi
Ding, Ming
Source :
Photonic Sensors; Jun2019, Vol. 9 Issue 2, p189-196, 8p
Publication Year :
2019

Abstract

We propose an in-situ method to calibrate the coil constants of the optical atomic magnetometer. This method is based on measuring the Larmor precession of spin polarized alkali metal atoms and has been demonstrated on a K-Rb hybrid atomic magnetometer. Oscillation fields of different frequencies are swept on the transverse coil. By extracting the resonance frequency through phase-frequency analysis of electron spin projection, the coil constants are calibrated to be 323.1 ± 0.28 nT/mA, 108 ± 0.04 nT/Ma, and 185.8 ± 1.03 nT/mA along the X, Y, and Z directions, respectively. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
16749251
Volume :
9
Issue :
2
Database :
Complementary Index
Journal :
Photonic Sensors
Publication Type :
Academic Journal
Accession number :
135840414
Full Text :
https://doi.org/10.1007/s13320-019-0530-4