Comparison of dosimetric parameters of small-field electron beams between Advanced Markus, Semiflex 3D, and Diode E responses.
- Source :
- Radiological Physics & Technology; Sep2020, Vol. 13 Issue 3, p296-305, 10p
- Publication Year :
- 2020
-
Abstract
- The usage of dosimetry of small fields in radiotherapy to measure radiation dose is difficult because of high-dose gradients, lateral electronic disequilibrium, and detector volume effects. In this study, three dosimeters namely, Markus, Semiflex 3D, and Diode E were tested using the Elekta-accelerator electron beams. The electron beam parameters, penumbra, and output factor were determined using these dosimeters for each field size and energy. According to the results, Diode E and Advanced Markus exhibited the greatest difference in Rq among the electron beam parameters. Furthermore, the greatest difference in penumbra was observed between Diode E and Advanced Markus for the field size of 3 cm<superscript>2</superscript> at 10 MeV. In terms of output factor, three dosimeters exhibited the greatest difference between Diode E and Advanced Markus for the field size of 3 cm<superscript>2</superscript> at 10 MeV. The findings indicate that the Semiflex 3D can be regarded as an appropriate dosimeter for electron small-field dosimetry. [ABSTRACT FROM AUTHOR]