Analytical method for the determination of 211At using an α-scintillation camera system and thin-layer chromatography.
- Source :
- Journal of Radioanalytical & Nuclear Chemistry; Oct2020, Vol. 326 Issue 1, p773-778, 6p
- Publication Year :
- 2020
-
Abstract
- <superscript>211</superscript>At is artificially produced for medical purposes. To prevent severe loss of short-lived <superscript>211</superscript>At, a rapid method analyzing both the radioactivity and all chemical forms was required for related fundamental researches. We propose a new method by means of an α-scintillation-camera system comprising of thin-layer chromatography and a high-sensitive CCD camera. The performance of the system was experimentally verified: low-radioactivity <superscript>211</superscript>At of 56–672 Bq was measured in a short-time of 1000 s and all chemical forms of <superscript>211</superscript>At were evaluated. The results reveal that the method enables us to analyze two features of <superscript>211</superscript>At with significantly short-time. [ABSTRACT FROM AUTHOR]
- Subjects :
- CHROMATOGRAPHIC analysis
SCINTILLATION counters
CAMERAS
RADIOACTIVITY
CCD cameras
Subjects
Details
- Language :
- English
- ISSN :
- 02365731
- Volume :
- 326
- Issue :
- 1
- Database :
- Complementary Index
- Journal :
- Journal of Radioanalytical & Nuclear Chemistry
- Publication Type :
- Academic Journal
- Accession number :
- 146303295
- Full Text :
- https://doi.org/10.1007/s10967-020-07353-9