The impact of different artificial disc heights during total cervical disc replacement: an in vitro biomechanical study.

Authors :
Wang, Xiao-Fei
Meng, Yang
Liu, Hao
Wang, Bei-Yu
Hong, Ying
Source :
Journal of Orthopaedic Surgery & Research; 1/6/2021, Vol. 16 Issue 1, p1-10, 10p
Publication Year :
2021

Abstract

Background: The principles of choosing an appropriate implant height remain controversial in total cervical disc replacement (TDR). By performing an in vitro biomechanical study and exploring the biomechanical impact of implant height on facet joint and motion function, the study aimed to offer valid proposals regarding implant height selection during TDR. Methods: A total of 6 fresh-frozen male cadaveric cervical spines (C2–C7) with 5 mm intervertebral disc height at C5/6 level were enrolled in the study. Specimens with the intact condition and with different height artificial discs were tested. Facet joint pressures and range of motion under each condition were recorded using a specialized machine. Results: The artificial disc heights that were involved in this study were 5 mm, 6 mm, and 7 mm. The range of motion decreased along with the increment of implant height, while facet joint pressure showed an opposite trend. Specimens with a 5 mm implant height could provide a similar range of motion (11.8° vs. 12.2° in flexion-extension, 8.7° vs. 9.0° in rotation, 7.9° vs. 8.2° in lateral bending) and facet joint pressure (27.8 psi vs. 25.2 psi in flexion, 59.7 psi vs. 58.9 psi in extension, 24.0 psi vs. 22.7 psi in rotation, 32.0 psi vs. 28.8 psi in lateral bending) compared with intact specimens. Facet joint pressure of specimens with 6 mm implant height (≥ 1 mm in height) increased during flexion at the C5–6 segment (30.4 psi vs. 25.2 psi, P = 0.076). However, specimens with 7 mm implant height (≥ 2 mm in height) showed a significant reduction in motion (9.5° vs. 12.2° in flexion-extension, P < 0.001) and increment of facet joint pressure at C5–6 segment (44.6 psi vs. 25.2 psi in flexion, 90.3 psi vs. 58.9 psi in extension, P < 0.0001) and adjacent segments. Conclusions: This study suggested that an appropriate artificial disc height can achieve near-normal biomechanical properties and is recommended. We should be very cautious when using artificial discs ≥ 1 mm in height compared to normal. However, implants ≥ 2 mm in height compared to normal significantly increased the facet joint pressure and decreased the range of motion; therefore, it should not be used in clinical practice. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
1749799X
Volume :
16
Issue :
1
Database :
Complementary Index
Journal :
Journal of Orthopaedic Surgery & Research
Publication Type :
Academic Journal
Accession number :
147948442
Full Text :
https://doi.org/10.1186/s13018-020-02157-9