Preliminary performance measurements for a streak camera with a large-format direct-coupled charge-coupled device readout.
- Source :
- Review of Scientific Instruments; Oct2004 Part I & II, Vol. 75 Issue 10, p4042-4044, 3p
- Publication Year :
- 2004
-
Abstract
- The University of Rochester’s Laboratory for Laser Energetics (Rochester, New York) is leading an effort to develop a modern, fully automated streak camera. Characterization of a prototype camera shows spatial resolution better than 20 lp/mm, temporal resolution of 12 ps, line-spread function of 40 μm (full width at half maximum) contrast transfer ratio of 60% at 10 lp/mm, system gain of 101 charge-coupled device electrons per photoelectron, and a dynamic range of 500 for a 2 ns window. [ABSTRACT FROM AUTHOR]