Measurement of urine albumin by liquid chromatography-isotope dilution tandem mass spectrometry and its application to value assignment of external quality assessment samples and certification of reference materials.

Authors :
Chen, Yizhao
Liu, Hong
Loh, Tze Ping
Liu, Qinde
Teo, Tang Lin
Lee, Tong Kooi
Sethi, Sunil Kumar
Source :
Clinical Chemistry & Laboratory Medicine; Apr2021, Vol. 59 Issue 4, p711-720, 10p
Publication Year :
2021

Abstract

Urine albumin is measured in clinical laboratories by immunoturbidimetry. However, large biases are observed among the different routine methods. To standardize the measurement of urine albumin, a reference measurement procedure (RMP) and urine albumin certified reference materials (CRMs) are needed. A candidate RMP for urine albumin based on liquid chromatography-isotope dilution tandem mass spectrometry (LC-IDMS/MS) using human serum albumin as calibration standard was developed. Isotope-labeled human albumin was used as internal standard. Urine samples were digested using trypsin and eight resulting "signature" peptides of albumin were quantified by LC-IDMS/MS. The candidate RMP was employed in value assignment of external quality assessment (EQA) samples and certification of urine albumin reference materials. The commutability of the developed CRMs was assessed against patient samples. The candidate RMP (recovery 101.5–103.2% and CV 1.2–3.3% at about 7–40 mg/L) met optimal performance goal. The lower limit of quantification was 0.03 mg/L as determined by signal-to-noise method. The EQA results from clinical laboratories using different immunoturbidimetric methods were generally comparable with assigned target values determined by the candidate RMP, with albumin concentrations ranging from 5 to 226 mg/L. Urine albumin reference materials (two levels) certified using the candidate RMP showed good commutability in a preliminary study. With optimal method precision and trueness, as well as comparability with routine methods, the developed RMP may be used for value assignment of EQA samples or certification of reference materials, which are important pillars in urine albumin method standardization. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
14346621
Volume :
59
Issue :
4
Database :
Complementary Index
Journal :
Clinical Chemistry & Laboratory Medicine
Publication Type :
Academic Journal
Accession number :
149221749
Full Text :
https://doi.org/10.1515/cclm-2020-0969