Comparison of SimoaTM and EllaTM to assess serum neurofilament‐light chain in multiple sclerosis.

Authors :
Gauthier, Audrey
Viel, Sébastien
Perret, Magali
Brocard, Guillaume
Casey, Romain
Lombard, Christine
Laurent‐Chabalier, Sabine
Debouverie, Marc
Edan, Gilles
Vukusic, Sandra
Lebrun‐Frénay, Christine
De Sèze, Jérôme
Laplaud, David Axel
Castelnovo, Giovanni
Gout, Olivier
Ruet, Aurélie
Moreau, Thibault
Casez, Olivier
Clavelou, Pierre
Berger, Eric
Source :
Annals of Clinical & Translational Neurology; May2021, Vol. 8 Issue 5, p1141-1150, 10p
Publication Year :
2021

Abstract

We compared SimoaTM and EllaTM immunoassays to assess serum neurofilament‐light chain levels in 203 multiple sclerosis patients from the OFSEP HD study. There was a strong correlation (ρ = 0.86, p < 0.0001) between both platforms. The EllaTM instrument overestimated values by 17%, but as the data were linear (p = 0.57), it was possible to apply a correction factor to EllaTM results. As for SimoaTM, serum neurofilament‐light chain levels measured by EllaTM were correlated with age and EDSS and were significantly higher in active multiple sclerosis, suggesting that these assays are equivalent and can be used in routine clinical practice. [ABSTRACT FROM AUTHOR]