Abstract

<bold>Background: </bold>High power shorter duration (HPSD) ablation may lead to safe and rapid lesion formation. However, the optimal radio frequency power to achieve the desired ablation index (AI) or lesion size index (LSI) is insubstantial. This analysis aimed to appraise the clinical safety and efficacy of HPSD guided by AI or LSI (HPSD-AI or LSI) in patients with atrial fibrillation (AF).<bold>Methods: </bold>The Medline, PubMed, Embase, Web of Science, and the Cochrane Library databases from inception to November 2020 were searched for studies comparing HPSD-AI or LSI and low power longer duration (LPLD) ablation.<bold>Results: </bold>Seven trials with 1013 patients were included in the analysis. The analyses verified that HPSD-AI or LSI revealed benefits of first-pass pulmonary vein isolation (PVI) (RR: 1.28; 95% CI: 1.05-1.56, P = 0.01) and acute pulmonary vein reconnection (PVR) (RR: 0.65; 95% CI: 0.48-0.88, P = 0.005) compared with LPLD. HPSD-AI or LSI showed higher freedom from atrial tachyarrhythmia (AT) (RR = 1.32, 95% CI: 1.14-1.53, P = 0.0002) in the subgroup analysis of studies with PVI ± (with or without additional ablation beyond PVI). HPSD-AI or LSI could short procedural time (WMD: -22.81; 95% CI, -35.03 to -10.60, P = 0.0003), ablation time (WMD: -10.80; 95% CI: -13.14 to -8.46, P < .00001), and fluoroscopy time (WMD: -7.71; 95% CI: -13.71 to -1.71, P = 0.01). Major complications and esophageal lesion in HPSD-AI or LSI group were no more than LDLP group (RR: 0.58; 95% CI: 0.20-1.69, P = 0.32) and (RR: 0.84; 95% CI: 0.43-1.61, P = 0.59).<bold>Conclusions: </bold>HPSD-AI or LSI was efficient for treating AF with shorting procedural, ablation, and fluoroscopy time, higher first-pass PVI, and reducing acute PVR and may increase freedom from AT for patients with additional ablation beyond PVI compared with LPLD. Moreover, complications and esophageal lesion were low and no different between two groups. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
08964327
Database :
Complementary Index
Journal :
Journal of Interventional Cardiology
Publication Type :
Academic Journal
Accession number :
150666220
Full Text :
https://doi.org/10.1155/2021/5591590