Back to Search Start Over

Validation of a Monte Carlo Modelling Based Dosimetry of Extraoral Photobiomodulation.

Authors :
Yaroslavsky, Anna N.
Juliano, Amy F.
Adnan, Ather
Selting, Wayne J.
Iorizzo, Tyler W.
Carroll, James D.
Sonis, Stephen T.
Duncan, Christine N.
London, Wendy B.
Treister, Nathaniel S.
Source :
Diagnostics (2075-4418); Dec2021, Vol. 11 Issue 12, p2207-2207, 1p
Publication Year :
2021

Abstract

An in vivo validation study was performed to confirm the accuracy of extraoral photobiomodulation therapy (PBMT) dosimetry determined by modelling. The Monte Carlo technique was utilized to calculate the fluence rate and absorbed power of light delivered through multi-layered tissue. Optical properties used during Monte Carlo simulations were taken from the literature. Morphological data of four study volunteers were acquired using magnetic resonance imaging (MRI) scans. Light emitting diode (LED) coupled to a power meter were utilized to measure transmitted power through each volunteer's cheek, in vivo. The transmitted power determined by Monte Carlo modelling was compared to the in vivo measurements to determine the accuracy of the simulations. Experimental and simulation results were in good agreement for all four subjects. The difference between the mean values of the measured transmission was within 12% from the respective transmission obtained using Monte Carlo simulations. The results of the study indicate that Monte Carlo modelling is a robust and reliable method for light dosimetry. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
20754418
Volume :
11
Issue :
12
Database :
Complementary Index
Journal :
Diagnostics (2075-4418)
Publication Type :
Academic Journal
Accession number :
154372120
Full Text :
https://doi.org/10.3390/diagnostics11122207