Portable fluorescence instrument for detecting membrane integrity in membrane bioreactor (MBR)

Authors :
Yu, Yang
Xin, Changchun
Liu, Yuxiang
Gao, Fei
Zhang, Lei
Jia, Hui
Wang, Jie
Source :
Frontiers of Environmental Science & Engineering; Feb2024, Vol. 18 Issue 2, p1-11, 11p
Publication Year :
2024

Abstract

This study proposed the design, fabrication, and assembly of membrane integrity detection instruments in membrane bioreactors (MBR) based on fluorescence spectroscopy. Based on the PARAFAC model, we found that the peak at 280/335 nm strengthened after membrane breakage. The peak at 340/430 nm reflected the sludge concentration in the MBR and reduced the influence of internal filtration effects on detection. Therefore, we determined that the dual-LED light source excitation detection system can detect tryptophan-like substances at 280 nm (T-peak) and humic acid at 340 nm (C-peak). T-peak was identified as the core index indicating membrane integrity. Moreover, the C-peak is the reference indicator factor for a sensitive response to changes in the sludge concentration. The portable fluorescence instrument exhibited high sensitivity and good feedback accuracy compared to particle counting and turbidity detection, where the log reduction value was greater than 3.5. This overcomes the disadvantage of false alarms in particle counters and is not affected by the position of the pump system. This portable instrument provides a flexible and highly sensitive method for the assessment of industrial membrane integrity. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
20952201
Volume :
18
Issue :
2
Database :
Complementary Index
Journal :
Frontiers of Environmental Science & Engineering
Publication Type :
Academic Journal
Accession number :
173981690
Full Text :
https://doi.org/10.1007/s11783-024-1783-8