Back to Search
Start Over
On-Site Calibration Procedure and Uncertainty Contributions on Air Temperature Sensors.
- Source :
- International Journal of Thermophysics; Jan2024, Vol. 45 Issue 1, p1-14, 14p
- Publication Year :
- 2024
-
Abstract
- Due to the increasing demand of climate change studies, traceability of measurements according to the International System of Units (SI) becomes fundamental to establish data quality and comparability in time. With the aim of guaranteeing the traceability, in October 2022, an on-site calibration of three thermometers operated by the Osservatorio meteorologico di Moncalieri was assessed. The calibration by comparison against traceable travelling standards was performed in a transportable thermal chamber manufactured by the Istituto Nazionale di Ricerca Metrologica (INRIM). Results showed that the main uncertainty contributions to the expanded uncertainty were the interpolation method, the chamber inhomogeneity and the repeatability of the sensors under calibration. This current calibration was compared with previous calibration in 2012 (on-site calibration) and in 2016 (laboratory calibration). The comparison analysis evidenced the drift effect on the thermometers and the importance of having an active calibration program to reduce this effect. Regarding the expanded uncertainty, both on-site calibrations presented the same order of magnitude and smaller than the correction values. This paper points out the advantage of performing an on-site calibration: since the calibration involves the same dataloggers and cables in the same environmental conditions, the calculated calibration curve represents more convincingly the real measurement conditions. [ABSTRACT FROM AUTHOR]
- Subjects :
- TEMPERATURE sensors
METRIC system
ATMOSPHERIC temperature
CALIBRATION
Subjects
Details
- Language :
- English
- ISSN :
- 0195928X
- Volume :
- 45
- Issue :
- 1
- Database :
- Complementary Index
- Journal :
- International Journal of Thermophysics
- Publication Type :
- Academic Journal
- Accession number :
- 174798885
- Full Text :
- https://doi.org/10.1007/s10765-023-03296-x