Back to Search Start Over

A Time-of-Flight and Radar Dataset of a neonatal Thorax Simulator with synchronized Reference Sensor Signals for respiratory Rate Detection.

Authors :
Gleichauf, Johanna
Herrmann, Sven
Niebler, Christine
Koelpin, Alexander
Source :
Scientific Data; 1/23/2024, Vol. 11 Issue 1, p1-8, 8p
Publication Year :
2024

Abstract

In this paper we present an open-source Time-of-Flight and radar dataset of a neonatal thorax simulator for the development of respiratory rate detection algorithms. As it is very difficult to gain recordings of (preterm) neonates and there is hardly any open-source data available, we built our own neonatal thorax simulator which simulates the movement of the thorax due to respiration. We recorded Time-of-Flight (ToF) and radar data at different respiratory rates in a range of 5 to 80 breaths per minute (BPM) and with varying upstroke heights. As gold standard a laser micrometer was used. The open-source data can be used to test new algorithms for non-contact respiratory rate detection. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
20524463
Volume :
11
Issue :
1
Database :
Complementary Index
Journal :
Scientific Data
Publication Type :
Academic Journal
Accession number :
174971389
Full Text :
https://doi.org/10.1038/s41597-024-02946-5