Back to Search Start Over

The Application of Optical Sensors with Built-in Anchor-like Cavities in the Detection of Hemoglobin Concentration.

Authors :
Jiang, Wen
Yan, Shubin
Su, Yiru
Wang, Chong
Wu, Taiquan
Cui, Yang
Zhu, Chuanhui
Zhang, Yi
Mu, Xiangyang
Gao, Guowang
Source :
Photonics; May2024, Vol. 11 Issue 5, p402, 14p
Publication Year :
2024

Abstract

This paper introduces a refractive index sensor based on Fano resonance, utilizing a metal–insulator–metal (MIM) waveguide structure with an Anchor-like cavity. This study utilizes the finite element method (FEM) for analyzing the propagation characteristics of the structure. The evaluation concentrated on assessing how the refractive index and the structure's geometric parameters affect its sensing characteristics. The designed structure demonstrates optimum performance, achieving a maximum sensitivity of 2440 nm/RIU and an FOM of 63. Given its high sensitivity, this nanoscale refractive index sensor is ideal for detecting hemoglobin concentrations in blood, and the sensor's sensitivity is 0.6 nm·g/L, aiding in clinical prevention and treatment. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
23046732
Volume :
11
Issue :
5
Database :
Complementary Index
Journal :
Photonics
Publication Type :
Academic Journal
Accession number :
177488411
Full Text :