Back to Search Start Over

Development of Portable Multi-fluorescence Detection System Using Indium Tin Oxide Heater for Loop-mediated Isothermal Amplification.

Authors :
Ryo Ishii
Sota Hirose
Shoji Yamamoto
Kazuhiro Morioka
Akihide Hemmi
Hizuru Nakajima
Source :
Sensors & Materials; 2024, Vol. 36 Issue 8, Part 1, p3189-3199, 11p
Publication Year :
2024

Abstract

We developed a portable multi-fluorescence detection system for loop-mediated isothermal amplification (LAMP) using an indium tin oxide (ITO) transparent glass heater by improving a previously developed device. By changing the optical design and the material of the heating part, the quantitative performance and the stability of temperature control were improved. We demonstrated the basic performance of the developed device for genetic tests using LAMP by conducting discrimination tests for human genome DNA. We also showed that quantitative analysis of human genomic DNA is possible at initial DNA concentrations of 2.11 × 10²– 2.11 × 10<superscript>4</superscript> ng/µL using the developed device. This genetic testing device is smaller and lighter (60 × 90 × 70 mm³, 178 g) than conventional testing devices, giving it considerable potential to facilitate rapid and efficient on-site genetic testing. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
09144935
Volume :
36
Issue :
8, Part 1
Database :
Complementary Index
Journal :
Sensors & Materials
Publication Type :
Academic Journal
Accession number :
178892285
Full Text :
https://doi.org/10.18494/SAM4635