Back to Search Start Over

Modeling a one-dimensional bremsstrahlung and neutron imaging array for use on Sandia's Z machine.

Authors :
Rochau, G.A.
Derzon, M.S.
Fehl, D.
Rochau, G.E.
Lazier, S.
Droemer, D.
Source :
Review of Scientific Instruments; Jan1999, Vol. 70 Issue 1, p549, 4p, 4 Diagrams, 4 Graphs
Publication Year :
1999

Abstract

Outlines the shielding design of the time-resolved bremsstrahlung and neutron camera diagnostic intended to image both bremsstrahlung radiation and the neutron emission from z-pinch target regions on the Z accelerator at Sandia National Laboratories. Array design featuring scintillator fibers separated by tungsten collimator plates; External shielding design.

Details

Language :
English
ISSN :
00346748
Volume :
70
Issue :
1
Database :
Complementary Index
Journal :
Review of Scientific Instruments
Publication Type :
Academic Journal
Accession number :
4270373
Full Text :