Back to Search Start Over

Energy and Timing Resolution Studies With Silicon Photomultipliers (SiPMs) and 4-Pixel SiPM Matrices for PET.

Authors :
Llosá, Gabriela
Belcari, Nicola
Bisogni, Maria Giuseppina
Collazuol, Gianmaria
Marcatili, Sara
Moehrs, Sascha
Morsani, Fabio
Piemonte, Claudio
del Guerra, Alberto
Source :
IEEE Transactions on Nuclear Science; Jun2009 Part 1 of 3, Vol. 56 Issue 3, p543-548, 6p, 9 Graphs
Publication Year :
2009

Abstract

A high performance detector head with matrices of silicon photomultipliers (SiPMs) as photodetectors is under development at the University of Pisa and INFN Pisa. The silicon photomultipliers fabricated at FBK-irst (Trento, Italy) are being evaluated for this purpose. Single SiPM detectors of 1 mm × 1 mm size and the first SiPM test matrices composed of four (2 × 2) pixel elements in a common substrate have been tested with LYSO crystals. The energy and timing resolution of single SiPMs have been evaluated, obtaining an energy resolution of 20% FWHM at 511 keV, and a coincidence timing resolution of 1.4 ns FWHM. In addition, the first tests performed with the SiPM matrices prove the feasibility of such devices. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00189499
Volume :
56
Issue :
3
Database :
Complementary Index
Journal :
IEEE Transactions on Nuclear Science
Publication Type :
Academic Journal
Accession number :
43153399
Full Text :
https://doi.org/10.1109/TNS.2009.2015877