Back to Search
Start Over
Characterization of Thermal Cross-talk in a γ-ray Microcalorimeter Array.
- Source :
- AIP Conference Proceedings; 12/16/2009, Vol. 1185 Issue 1, p34-37, 4p
- Publication Year :
- 2009
-
Abstract
- We present experimental data describing cross-talk within an array of gamma-ray microcalorimeters during gamma-ray irradiation. The microcalorimeters consist of Mo/Cu transition-edge sensors (TESs) with attached Sn absorbers. We observe both thermal and electrical cross-talk with peak cross-talk amplitudes as large as 0.4%. We have developed an analytical model for thermal cross-talk and make a preliminary comparison to data. Cross-talk must be understood and minimized for high resolution spectroscopy at high input count rates. [ABSTRACT FROM AUTHOR]
- Subjects :
- CALORIMETERS
SPECTRUM analysis
GAMMA rays
IRRADIATION
HIGH resolution spectroscopy
Subjects
Details
- Language :
- English
- ISSN :
- 0094243X
- Volume :
- 1185
- Issue :
- 1
- Database :
- Complementary Index
- Journal :
- AIP Conference Proceedings
- Publication Type :
- Conference
- Accession number :
- 47200660
- Full Text :
- https://doi.org/10.1063/1.3292349