Characterization of Thermal Cross-talk in a γ-ray Microcalorimeter Array.

Authors :
Jethava, N.
Ullom, J. N.
Bennett, D. A.
Irwin, K. D.
Horansky, R. D.
Beall, J. A.
Hilton, G. C.
Vale, L. R.
Hoover, A.
Bacrania, M. K.
Rabin, M. W.
Source :
AIP Conference Proceedings; 12/16/2009, Vol. 1185 Issue 1, p34-37, 4p
Publication Year :
2009

Abstract

We present experimental data describing cross-talk within an array of gamma-ray microcalorimeters during gamma-ray irradiation. The microcalorimeters consist of Mo/Cu transition-edge sensors (TESs) with attached Sn absorbers. We observe both thermal and electrical cross-talk with peak cross-talk amplitudes as large as 0.4%. We have developed an analytical model for thermal cross-talk and make a preliminary comparison to data. Cross-talk must be understood and minimized for high resolution spectroscopy at high input count rates. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
0094243X
Volume :
1185
Issue :
1
Database :
Complementary Index
Journal :
AIP Conference Proceedings
Publication Type :
Conference
Accession number :
47200660
Full Text :
https://doi.org/10.1063/1.3292349