Suitability and calibration of a rebound tonometer to measure IOP in rabbit and pig eyes.

Authors :
Löbler, Marian
Rehmer, Annelie
Guthoff, Rudolf
Martin, Heiner
Sternberg, Katrin
Stachs, Oliver
Source :
Veterinary Ophthalmology; Jan/Feb2011, Vol. 14 Issue 1, p66-68, 3p
Publication Year :
2011

Abstract

To utilize the Icare tonometer TAO1 for intraocular pressure (IOP) determination in experimental animals. To calculate true IOP calibration functions for rabbit and porcine eyes. Enucleated eyes of 3-year-old healthy experimental rabbits (New Zealand white), and healthy 1 year old experimental pigs (Deutsche Landrasse) were used for the determination of IOP. Manometric (Geuder GmbH, Heidelberg/Germany) and rebound tonometry (Icare tonometer TAO1, Icare, Helsinki/Finland) were used to record IOP in enucleated animal eyes (rabbit n = 2, pig n = 3). The Icare tonometer TAO1 measurements underestimated true IOP by 37-60% in rabbit eyes and 17-63% in porcine eyes. IOP values obtained by both rebound and manometric tonometry for rabbit and porcine eyes followed a linear regression curve. Linear functions were calculated to correct the Icare tonometer TAO1 measurements to true IOP for both rabbit ( p = 1.4244 p + 4.2421) and porcine eyes ( p = 1.0799 p + 5.8557). The Icare tonometer TAO1 can be utilized for IOP determination in rabbit and porcine eyes when measured values are corrected with the appropriate linear function. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
14635216
Volume :
14
Issue :
1
Database :
Complementary Index
Journal :
Veterinary Ophthalmology
Publication Type :
Academic Journal
Accession number :
56942956
Full Text :
https://doi.org/10.1111/j.1463-5224.2010.00794.x