Back to Search Start Over

Automation of Direct Josephson Voltage Standard Comparison and an Application.

Authors :
Tang, Yi-hua
Source :
IEEE Transactions on Instrumentation & Measurement; Jun2011, Vol. 60 Issue 7, p2404-2408, 5p
Publication Year :
2011

Abstract

A direct comparison between two Josephson voltage standard (JVS) systems has been used in the Bureau International des Poids et Mesures key comparisons or regional supplementary comparisons to ensure the uniformity of the representation of the volt based on K{J - 90} with a relative uncertainty of a few parts in 10^10 at a nominal voltage of 10 V. Most direct JVS comparisons are manually operated. Hardware and protocol for an automatic direct JVS comparison have been developed to verify the performance of a JVS while retaining the same level of uncertainty attained by a manual comparison. The automatic JVS direct comparison significantly improves the efficiency of the data acquisition. A case study examining the offset related to the digital voltmeter polarity in an automatic direct JVS comparison is presented. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00189456
Volume :
60
Issue :
7
Database :
Complementary Index
Journal :
IEEE Transactions on Instrumentation & Measurement
Publication Type :
Academic Journal
Accession number :
61129109
Full Text :
https://doi.org/10.1109/TIM.2011.2108611