Back to Search Start Over

Uncertainty Evaluations of an AC Shunt Calibration System With a Load Effect Reduction Circuit.

Authors :
Kon, Saytaro
Yamada, Tatsuji
Source :
IEEE Transactions on Instrumentation & Measurement; Jun2011, Vol. 60 Issue 7, p2286-2291, 6p
Publication Year :
2011

Abstract

An ac shunt calibration system at the National Metrology Institute of Japan (NMIJ) has been developed. The uncertainties of the system were estimated 3.4 \mu\Omega/\Omega for in-phase and 8.6 \mu\rad for quadrature at 55 Hz/5 A. Meanwhile, the enhancement of the system is planned up to 10 kHz/10 A. For this goal, the accuracy and high-frequency performance of the buffer amplifier, which have the largest influence on the uncertainties, have been improved by developing a load effect reduction (LER) circuit. The LER circuit cancels out the effect of output impedance of the buffer amplifier. The effect of the circuit on the buffer amplifier allows the uncertainties of the ac shunt calibration system to be reduced into 2.6 \mu\Omega/\Omega and 8.2 \mu\rad at power frequencies. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00189456
Volume :
60
Issue :
7
Database :
Complementary Index
Journal :
IEEE Transactions on Instrumentation & Measurement
Publication Type :
Academic Journal
Accession number :
61129169
Full Text :
https://doi.org/10.1109/TIM.2010.2103414