Blood drawn through valved catheter hub connectors carries a significant risk of contamination.

Authors :
Sherertz, R.
Karchmer, T.
Palavecino, E.
Bischoff, W.
Source :
European Journal of Clinical Microbiology & Infectious Diseases; Dec2011, Vol. 30 Issue 12, p1571-1577, 7p
Publication Year :
2011

Abstract

Infection Control became concerned when bloodstream infection (BSI) rates increased after implementing a needleless valved hub connector. During a 21-month period three different needleless catheter hub connectors were evaluated by quantitatively culturing blood drawn through hub connectors that would have ordinarily been discarded (DBC). DBC drawn through Clearlink™ catheter hub connectors were found to be twice as likely to be positive as DBC drawn through Clave® or Q-syte™ hub connectors ( P < 0.04). DBC grew pathogens 46% of the time and skin organisms 54% of the time. Patients with positive DBC were three times more likely to meet Centers for Disease Control (CDC) BSI criteria by DBC cultures than by physician-ordered blood cultures (CBC; P < 0.001). For patients growing pathogens in DBC, 64% had no CBC drawn, the average temperature was lower than for patients with pathogens in CBC (99.3 ± 1.5 ve 100.6 ± 1.9, P = 0.015), and 92% of discharged patients (11 out of 12) were not treated with an antibiotic active against the DBC pathogen. Drawing BC through a catheter hub connector carries a risk of false-positives that could increase BSI rates by up to 3-fold. Further work is necessary to evaluate this concern. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
09349723
Volume :
30
Issue :
12
Database :
Complementary Index
Journal :
European Journal of Clinical Microbiology & Infectious Diseases
Publication Type :
Academic Journal
Accession number :
67079295
Full Text :
https://doi.org/10.1007/s10096-011-1262-6