Dosimetry Study for a New in Vivo XRF Bone Lead Measurement System.

Authors :
Hou, Dejia
Luo, Xianqing
Tan, Zuojun
Wang, Haiying
Yi, Weisong
Wang, Xianfeng
Source :
2013 Fifth International Conference on Measuring Technology & Mechatronics Automation; 2013, p485-488, 4p
Publication Year :
2013

Abstract

This paper discusses a new in vivo XRF bone lead detection system and introduces how to study the radioactive activity, exposure rate, absorb doses, equivalent dose and effective dose of the system using the long life source of 238Pu and analyzes the dose safety about the system. [ABSTRACT FROM PUBLISHER]