Back to Search
Start Over
A cw AgGaS[sub 2] difference frequency spectrometer with diode lasers as pump sources.
- Source :
- Applied Physics B: Lasers & Optics; 1997, Vol. 64 Issue 5, p521, 4p
- Publication Year :
- 1997
-
Abstract
- A tunable high resolution cw difference frequency spectrometer applying the non-linear crystal AgGaS[sub 2] and two diode lasers as pump sources and experiments with this setup are presented. In contrast to the majority of known applications of the difference frequency mixing with solid state or gas lasers, we apply two single mode diode lasers with emission wavelengths of 690 nm and 805 nm, respectively. By fixing the emission wavelength of one diode laser and tuning the wavelength of the second laser, by changing temperature or excitation current, we can cover typically 0.87 cm[sup -1] (FWHM) in good agreement with recently published data and theory. With an input power of 20 mW and 6 mW we achieved an output power in the nW-range. To demonstrate the capabilities of the spectrometer we scanned CO and OCS absorption lines near 2107 cm[sup -1] . From these experiments we deduce an overall signal-to-noise ratio of 1000:1 and a spectral resolution better than 30 MHz. With such parameters a trace gas detection of CO at sub-ppm level will be possible. [ABSTRACT FROM AUTHOR]
- Subjects :
- SPECTROMETERS
LASERS
DIODES
Subjects
Details
- Language :
- English
- ISSN :
- 09462171
- Volume :
- 64
- Issue :
- 5
- Database :
- Complementary Index
- Journal :
- Applied Physics B: Lasers & Optics
- Publication Type :
- Academic Journal
- Accession number :
- 8825559
- Full Text :
- https://doi.org/10.1007/s003400050209