Ion source and injection line for high intensity medical cyclotron.

Authors :
XianLu Jia
Fengping Guan
Hongjuan Yao
TianJue Zhang
Jianjun Yang
Guofang Song
Tao Ge
Jiuchang Qin
Source :
Review of Scientific Instruments; 2014, Vol. 85 Issue 2, p1-3, 3p
Publication Year :
2014

Abstract

A 14 MeV high intensity compact cyclotron, CYCIAE-14, was built at China Institute of Atomic Energy (CIAE). An injection system based on the external H- ion source was used on CYCIAE- 14 so as to provide high intensity beam, while most positron emission tomography cyclotrons adopt internal ion source. A beam intensity of 100 μA/14 MeV was extracted from the cyclotron with a small multi-cusp H- ion source (CIAE-CH-I type) and a short injection line, which the H- ion source of 3 mA/25 keV H- beam with emittance of 0.3π mmmrad and the injection line of with only 1.2 m from the extraction of ion source to the medial plane of the cyclotron. To increase the extracted beam intensity of the cyclotron, a new ion source (CIAE-CH-II type) of 9.1 mA was used, with maximum of 500 μA was achieved from the cyclotron. The design and test results of the ion source and injection line optimized for high intensity acceleration will be given in this paper. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00346748
Volume :
85
Issue :
2
Database :
Complementary Index
Journal :
Review of Scientific Instruments
Publication Type :
Academic Journal
Accession number :
97144198
Full Text :
https://doi.org/10.1063/1.4829739