Back to Search Start Over

Resonance response of scanning force microscopy cantilevers.

Authors :
Chen, G. Y.
Warmack, R. J.
Thundat, T.
Allison, D. P.
Huang, A.
Source :
Review of Scientific Instruments; Aug1994, Vol. 65 Issue 8, p2532, 6p
Publication Year :
1994

Abstract

A variational method is used to calculate the deflection and the fundamental and harmonic resonance frequencies of commercial V-shaped and rectangular atomic force microscopy cantilevers. The effective mass of V-shaped cantilevers is roughly half that calculated for the equivalent rectangular cantilevers. Damping by environmental gases, including air, nitrogen, argon, and helium, affects the frequency of maximum response and to a much greater degree the quality factor Q. Helium has the lowest viscosity, resulting in the highest Q, and thus provides the best sensitivity in noncontact force microscopy. Damping in liquids is dominated by an increase in effective mass of the cantilever due to an added mass of the liquid being dragged with that cantilever. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
00346748
Volume :
65
Issue :
8
Database :
Complementary Index
Journal :
Review of Scientific Instruments
Publication Type :
Academic Journal
Accession number :
9781885
Full Text :
https://doi.org/10.1063/1.1144647