Back to Search
Start Over
Accelerometer-Based Navigation Is as Accurate as Optical Computer Navigation in Restoring the Joint Line and Mechanical Axis After Total Knee Arthroplasty: A Prospective Matched Study.
- Source :
- Journal of Arthroplasty; Jan2016, Vol. 31 Issue 1, p92-97, 6p
- Publication Year :
- 2016
-
Abstract
- The Zimmer iASSIST system is a novel accelerometer-based navigation system for TKA. 76 patients (76 knees) were prospectively matched for age, BMI, gender, diagnosis, and pre-operative scores, and underwent TKA using the iASSIST (n=38) or optical CAS (n=38). There were no significant differences in clinical outcomes or satisfaction rates at six months post-operatively (P>0.05). Mechanical axis was 1.8±1.3° in the iASSIST cohort versus 2.1±1.6° in the CAS cohort (P=0.543). There were no significant differences in number of outliers for mechanical axis (P=1.000), coronal femoral-component angle (P=0.693), coronal tibial-component angle (P=0.204) or joint line deviation (P=1.000). The duration of surgery was significantly longer in the CAS group (P<0.001), while the added cost of accelerometer-based navigation was approximately $1000 per operation. [ABSTRACT FROM AUTHOR]
Details
- Language :
- English
- ISSN :
- 08835403
- Volume :
- 31
- Issue :
- 1
- Database :
- Supplemental Index
- Journal :
- Journal of Arthroplasty
- Publication Type :
- Academic Journal
- Accession number :
- 111636582
- Full Text :
- https://doi.org/10.1016/j.arth.2015.06.048