Back to Search Start Over

Reproducibility of T1ρ and T2 quantification in a multi-vendor multi-site study.

Authors :
Lartey, R.
Nanavati, A.
Kim, J.
Li, M.
Xu, K.
Nakamura, K.
Shin, W.
Winalski, C.S.
Obuchowski, N.
Bahroos, E.
Link, T.M.
Hardy, P.A.
Peng, Q.
Liu, K.
Fung, M.
Wu, C.
Li, X.
Lartey, Richard
Nanavati, Andreas
Kim, Jeehun
Source :
Osteoarthritis & Cartilage; Feb2023, Vol. 31 Issue 2, p249-257, 9p
Publication Year :
2023

Abstract

<bold>Objective: </bold>To evaluate the multi-vendor multi-site reproducibility of two-dimensional (2D) multi-echo spin-echo (MESE) T2 mapping (product sequences); and to evaluate the longitudinal reproducibility of three-dimensional (3D) magnetization-prepared angle-modulated partitioned k-space spoiled gradient echo snapshots (MAPSS) T1ρ and T2 mapping (research sequences), and 2D MESE T2 mapping, separated by 6 months, in a multi-vendor multi-site setting.<bold>Methods: </bold>Phantoms and volunteers (n = 5 from each site, n = 20 in total) were scanned on four 3 T magnetic resonance (MR) systems from four sites and three vendors (Siemens, General Electric, and Phillips). Two traveling volunteers (3 knees) scanned at all 4 sites at baseline and 6-month follow-up. Data was transferred to one site for centralized processing. Coefficients of variation (CVs) were calculated to evaluate reproducibility.<bold>Results: </bold>For baseline 2D MESE T2 measures, average CV were 0.37-2.45% (intra-site) and 5.96% (inter-site) for phantoms, and 3.15-8.49% (intra-site) and 14.16% (inter-site) for volunteers. For longitudinal phantom data, intra-site CVs were 1.42-3.48% for 3D MAPSS T1ρ, 1.77-3.56% for 3D MAPSS T2, and 1.02-2.54% for 2D MESE T2. For the longitudinal volunteer data, the intra-site CVs were 2.60-4.86% for 3D MAPSS T1ρ, 3.33-7.25% for 3D MAPSS T2, and 3.11-8.77% for 2D MESE T2.<bold>Conclusion: </bold>This study demonstrated excellent intra-site reproducibility of 2D MESE T2 imaging, while its inter-site variation was slightly higher than 3D MAPSS T2 imaging (10.06% as previously reported). This study also showed excellent reproducibility of longitudinal T1ρ and T2 cartilage quantification, in a multi-vendor multi-site setting for both product 2D MESE T2 and 3D MAPSS T1p/T2 research sequences. [ABSTRACT FROM AUTHOR]

Details

Language :
English
ISSN :
10634584
Volume :
31
Issue :
2
Database :
Supplemental Index
Journal :
Osteoarthritis & Cartilage
Publication Type :
Academic Journal
Accession number :
161527662
Full Text :
https://doi.org/10.1016/j.joca.2022.10.017