Back to Search
Start Over
Comparative Performance of Color-Measuring Instruments; Second Report
- Source :
- Applied Optics; June 1974, Vol. 13 Issue: 6 p1510-1518, 9p
- Publication Year :
- 1974
-
Abstract
- The comparative performance of twelve different late-model color-measuring instruments was studied for precision (short-term repeatability) and conformance of color measurement and color-difference measurement. The instruments tested included four true and two abridged spectrophotometers and six colorimeters, all being integrating-sphere instruments except three 45°/0° colorimeters. All the colorimeters were individually calibrated for close conformance to CIE coordinates. Forty-four samples taken from the group studied in the previous paper were measured three times on each instrument, by three different well-trained operators. Again, a GE spectrophotometer was considered the reference instrument for conformance studies. Overall, the GE spectrophotometer, the Hunterlab D25D colorimeter, and the Koll-morgen KCS-18 and KCS-40 abridged spectrophotometers demonstrated the best performance. Statistical treatment of the data has been improved vastly since the earlier paper, and some recalculated results are included.