Back to Search Start Over

High‐resolution 3D proton spectroscopic imaging of the human brain at 3 T: SNR issues and application for anatomy‐matched voxel sizes

Authors :
Gruber, Stephan
Mlynárik, Vladimir
Moser, Ewald
Source :
Magnetic Resonance in Medicine; February 2003, Vol. 49 Issue: 2 p299-306, 8p
Publication Year :
2003

Abstract

In a systematic study on the interdependence of linewidth, signal‐to‐noise ratio (SNR), and spatial resolution in 3D proton spectroscopic imaging (1H‐SI) at 3 T, we demonstrate reduced linewidths with increased spatial resolution due to reduced magnetic inhomogeneity within the brain. High‐precision quantitative data (0.75–0.094 cm3) were obtained for all resolutions, enabling the creation of metabolic maps that display details such as the ventricles, sulci, and gyri. High‐resolution 1H‐SI allows differences in metabolic ratios to be estimated for anatomically defined regions in gray (GM) and white matter (WM). Seven distinct regions in a healthy brain were anatomically segmented and their metabolic ratios were compared quantitatively. Data from a tumor patient are also presented to demonstrate potential clinical applications. Because of the high resolution, the metabolite ratios could be determined for distinct pathologic regions within the tumor and its surroundings. The method was additionally applied to a patient with patchy Pelizaeus Merzbacher disease (PMD), and compared to single‐voxel spectroscopy performed in the same session. High‐resolution SI data were demonstrated in our study to allow the direct matching of anatomic and metabolic images. This may enhance the clinical value of 1H‐SI. Magn Reson Med 49:299–306, 2003. © 2003 Wiley‐Liss, Inc.

Details

Language :
English
ISSN :
07403194 and 15222594
Volume :
49
Issue :
2
Database :
Supplemental Index
Journal :
Magnetic Resonance in Medicine
Publication Type :
Periodical
Accession number :
ejs24608348
Full Text :
https://doi.org/10.1002/mrm.10377