Back to Search Start Over

A new two-dimensional angular correlation of annihilation radiation apparatus using position-sensitive photomultiplier tubes

Authors :
Nagai, Yasuyoshi
Saito, Haruo
Iwata, Tetsuya
Nagashima, Yasuyuki
Hyodo, Toshio
Uchida, Hiroshi
Omura, Tomohide
Source :
Nuclear Instruments and Methods in Physics Research Section A; August 1996, Vol. 378 Issue: 3 p629-632, 4p
Publication Year :
1996

Abstract

A new two-dimensional angular correlation of annihilation radiation apparatus is described. Position-sensitive photomultiplier tubes coupled with two-dimensional arrays of small BGO scintillator blocks make simple and compact position-sensitive γ-ray detectors. With a sample-detector distance of 5 m, an angular resolution of 1.1 mrad FWHM and a coincidence count rate of ∼2.4 cps per mCi are obtained. Its performance is demonstrated by the result of a test measurement for KI crystal in which non-localized positronium exists at low temperatures.

Details

Language :
English
ISSN :
01689002
Volume :
378
Issue :
3
Database :
Supplemental Index
Journal :
Nuclear Instruments and Methods in Physics Research Section A
Publication Type :
Periodical
Accession number :
ejs2968113
Full Text :
https://doi.org/10.1016/0168-9002(96)00424-X