Back to Search Start Over

Decontamination of laryngoscopes in The Netherlands.

Authors :
Bucx, M J
Dankert, J
Beenhakker, M M
Harrison, T E
Source :
BJA: British Journal of Anaesthesia; January 2001, Vol. 86 Issue: 1 p99-102, 4p
Publication Year :
2001

Abstract

In this study the decontamination procedures of laryngoscopes in Dutch hospitals are described, based on a structured telephone questionnaire. There were substantial differences between decontamination procedures in Dutch hospitals and the standards of the APIC (Association of Professionals in Infection Control and Epidemiology), CDC (Centers of Disease Control) and ASA (American Society of Anesthesiology) were met in full in 19.4% of the hospitals. The standards of manual decontamination, used in 78% of the 139 hospitals, were particularly disappointing; manual cleaning was considered inadequate in 22.9% of these hospitals and manual disinfection did not meet the standards of the APIC, CDC or ASA in any of these hospitals. Decontamination by instrument cleaning machines as a standard procedure was used in 30 (22%) hospitals. In three of these hospitals the blades were subsequently sterilized. We suggest adherence to the infection control guidelines of the CDC, APIC and ASA, until the safety of less conservative infection control practices are demonstrated.

Details

Language :
English
ISSN :
00070912 and 14716771
Volume :
86
Issue :
1
Database :
Supplemental Index
Journal :
BJA: British Journal of Anaesthesia
Publication Type :
Periodical
Accession number :
ejs35881106
Full Text :