Back to Search Start Over

A highly sensitive electrochemical biosensor based on AuNP-modified gold electrodes for selective determination of serum levels of crosslaps

Authors :
Khashayar, Patricia
Amoabediny, Ghassem
Larijani, Bagher
Hosseini, Morteza
Verplancke, Rik
Keersmaecker, Michel
Adriaens, Annemie
Goemaere, Stefan
Fiers, Tom
Vanfleteren, Jan
Source :
3 Biotech; October 2017, Vol. 7 Issue: 5 p1-10, 10p
Publication Year :
2017

Abstract

This article explains a step-wise protocol to develop an electrochemical sensor to quantify serum levels of C-telopeptide (CTX) crosslinks also known as crosslaps in a matter of minutes and with high level of accuracy. The technique needs only one-step (incubation) and can thus be used for point of care screening. Due to the excellent electrical properties of the as-prepared immunosensor, CTX levels were successfully measured from 1 to 1000 pg/mL. This is while the normal reference of the marker is 50–450 pg/mL, suggesting that the sensor can acceptably detect CTX. The results also showed a good correlation with ECLIA in measuring serum levels of CTX.

Details

Language :
English
ISSN :
2190572X and 21905738
Volume :
7
Issue :
5
Database :
Supplemental Index
Journal :
3 Biotech
Publication Type :
Periodical
Accession number :
ejs43157208
Full Text :
https://doi.org/10.1007/s13205-017-0919-y