- Purpose To study the correlation among cataract grading system by Scheimpflug imaging (Oculus Pentacam®Nucleus Grading System [PNS]), effective phaco time (EPT), and surgical time (ST) using peristaltic and venturi pumps on the Whitestar Signature Phaco System (Abbott Medical Optics, Inc., Santa Ana, California, USA).Methods Patients scheduled to have bimanual microincision phacoemulsification by the same surgeon (T.F.) had Pentacam®Scheimpflug assessment using PNS and were randomly distributed into 2 groups. In group 1, surgery was performed using peristaltic pump, and in group 2, using venturi pump. Parameters assessed in both groups included EPT and ST.Results Sixty eyes that had uneventful clear cornea phacoemulsification with implantation of foldable intraocular lens (IOL) were evaluated, 30 in each group. The EPT was statistically superior with higher nuclear density grades for both groups 1 and 2 (group 1 Spearman R = 0.739, p<0.001; group 2 Spearman R = 0.924, p<0.001). The ST statistically increased with nucleus density for both groups 1 and 2 (group 1 Spearman R = 0.620, p<0.001; group 2 Spearman R = 0.643, p<0.001). Group 2 presented statistically significant reduction on median ST compared with group 1 for nucleus grading 2, 3, 4, and 5.Conclusions The PNS is an objective and reproducible method of grading nuclear cataract density and had a correlation with EPT. Peristaltic and venturi pumps in Signature Phaco System were similar in EPT but venturi pump reduced the median ST in nucleus with higher PNS, without repercussion on complications or anterior chamber stability.
Details
- Language :
- English
- ISSN :
- 11206721 and 17246016
- Volume :
- 23
- Issue :
- 6
- Database :
- Supplemental Index
- Journal :
- European Journal of Ophthalmology
- Publication Type :
- Periodical
- Accession number :
- ejs44949307
- Full Text :
- https://doi.org/10.5301/ejo.5000293