Back to Search
Start Over
Characterization of the Rotating Display
- Source :
- Investigative Radiology; September 1988, Vol. 23 Issue: 9 p692-694, 3p
- Publication Year :
- 1988
-
Abstract
- The rotating display is a useful method for reviewing single photon emission computed tomography (SPECT) data. This study evaluated the requirements for a subjectively pleasing and useful implementation of this technique. Twelve SPECT data sets were modified and viewed by several observers who recorded the minimum framing rates for apparent smooth rotation, 3D effect, effects of image size, and other parameters. The results showed that a minimum of 16 frames was needed for a useful display. Smaller image sizes and more frames were preferred. The recommended minimal framing rate for a 64-frame study is 16–17 frames per second and for a 32-frame study, 12–13 frames per second. Other enhancements also were useful.
Details
- Language :
- English
- ISSN :
- 00209996 and 15360210
- Volume :
- 23
- Issue :
- 9
- Database :
- Supplemental Index
- Journal :
- Investigative Radiology
- Publication Type :
- Periodical
- Accession number :
- ejs49114165