Back to Search Start Over

Sensitive Detection of Caffeic Acid and Rutin viathe Enhanced Anodic Electrochemiluminescence Signal of Luminol

Authors :
Xia, Tianlai
Gao, Yuan
Zhang, Ling
Wang, Xinyu
Pan, Guangxing
Wang, Zhenyuan
Han, Shuang
Ma, Xing
Zhao, Weiwei
Zhang, Jiaheng
Source :
Analytical Sciences; March 2020, Vol. 36 Issue: 3 p311-316, 6p
Publication Year :
2020

Abstract

The electrooxidation of phenolic groups of caffeic acid and rutin promote anodic electrochemiluminescence (ECL) luminol substantially. A sensitive, and cost-effective ECL method has thus been developed to detect caffeic acid, ranging from 0.1 to 5.0 μM, with a detection limit of 0.1 μM and rutin ranging from 0.2 to 25 μM with a detection limit of 0.12 μM. Contrarily, phenolic compounds quench the weak cathodic ECL of luminol. Both of anodic and cathodic ECL mechanisms of luminol in the presence of phenolic compounds are analyzed. The method based on the boomed anodic ECL of luminol is comparable to those based on Ru(bpy)3 2+ and S2O8 2–/O2 systems. A lower onset potential and price than the other ECL reagents would realize its widely applications in the detection of phenolic compounds in food and medicine.

Details

Language :
English
ISSN :
09106340 and 13482246
Volume :
36
Issue :
3
Database :
Supplemental Index
Journal :
Analytical Sciences
Publication Type :
Periodical
Accession number :
ejs60272984
Full Text :
https://doi.org/10.2116/analsci.19P274