Wireless Stimulation of Motor Cortex Through a Collagen Dura Substitute Using an Ultra-Thin Implant Fabricated on Parylene/PDMS

Authors :
Benbuk, Abed
Gulick, Daniel
Moniz-Garcia, Diogo
Liu, Shiyi
Quinones-Hinojosa, Alfredo
Christen, Jennifer Blain
Source :
IEEE Transactions on Biomedical Circuits and Systems; 2024, Vol. 18 Issue: 2 p334-346, 13p
Publication Year :
2024

Abstract

We present the design, fabrication, and in vivo testing of an ultra-thin (100 <inline-formula><tex-math notation="LaTeX">$\,\mathrm{\mu }\mathrm{m}$</tex-math></inline-formula>) wireless and battery-free implant for stimulation of the brain's cortex. The implant is fabricated on a flexible and transparent parylene/PDMS substrate, and it is miniaturized to dimensions of 15.6 × 6.6 mm<inline-formula><tex-math notation="LaTeX">$^{2}$</tex-math></inline-formula>. The frequency and pulse width of the monophasic voltage pulses are determined through On-Off keying (OOK) modulation of a wireless transmission at 2.45 GHz. Furthermore, the implant triggered a motor response in vivo when tested in 6 rodents. Limb response was observed by wireless stimulation of the brain's motor cortex through an FDA-approved collagen dura substitute that was placed on the dura in the craniotomy site, with no direct contact between the implant's electrodes and the brain's cortical surface. Therefore, the wireless stimulation method reported herein enables the concept of an e-dura substitute, where wireless electronics can be integrated onto a conventional dura substitute to augment its therapeutic function and administer any desired stimulation protocol without the need for post-surgical intervention for battery replacement or reprogramming stimulation parameters.

Details

Language :
English
ISSN :
19324545
Volume :
18
Issue :
2
Database :
Supplemental Index
Journal :
IEEE Transactions on Biomedical Circuits and Systems
Publication Type :
Periodical
Accession number :
ejs66011473
Full Text :
https://doi.org/10.1109/TBCAS.2023.3329447