Back to Search Start Over

Comparing Central Corneal Thickness Using Ultrasound And Anterior Segment Optical Coherence Tomography Pachymetry In Adults Attending A Private Eye Clinic In Abuja

Authors :
Okudo, Adaora C
Babalola, Olufemi E
Source :
Nigerian Medical Journal; Vol. 63 No. 1 (2022); 71-76
Publication Year :
2023
Publisher :
A Publication of the Nigerian Medical Association, 2023.

Abstract

Background: Central Corneal Thickness (CCT) measurement is useful in the management of glaucoma, ocular hypertension, corneal lesions and kerato-refractive surgeries. The study aims to compare the CCT measurements between Ultrasound Sonography (USS) and Optical Coherence Tomography (OCT), to analyze correlation and agreement between these instruments as well as repeatability of each instrument. Methodology: A cross sectional comparative study carried in 100 eyes of 50 patients attending Rachel Eye Center in Abuja from January to March 2021. The CCT were taken using the USS and OCT. CCT was measured using the Pachscan ultrasound and the Optovue machine utilised for the OCT technique. Data was analysed using SPSS version 20 using Paired Sample t Test, Pearson’s correlation, Interclass Correlation and Bland Altman Methods. Results: Patients were aged between 18 and 79 (mean age of 39.1), 72 males and 28 females. The mean CCT was 537.36 ±33.26 and 510.94 ±33.13 for USS and OCT respectively with a mean difference of (26.42±9.53 p

Details

Language :
English
ISSN :
03001652 and 2229774X
Database :
OpenAIRE
Journal :
Nigerian Medical Journal
Accession number :