Experimental characterization and test-beam results of MACACO III Compton camera
- Source :
- Digital.CSIC. Repositorio Institucional del CSIC, instname
- Publication Year :
- 2021
- Publisher :
- Institute of Electrical and Electronics Engineers, 2021.
-
Abstract
- The IRIS group at IFIC-Valencia is developing a Compton camera prototype with the aim of applying it in hadron therapy treatment monitoring. Recently, a third version of the prototype MACACO (Medical Applications CompAct COmpton camera) has been built. The system is composed of three Lanthanum (III) bromide scintillator crystals coupled to silicon photomultipliers. To improve its performance for the final application, several detectors are tested, two different silicon photomultipliers (25 and 50 um) have been chosen as possible candidates. The 25 up photodetector provided better performance in therms of dynamic range, energy resolution (5.2/ FWHM at 511 keV) and stability with temperature variations. MACACO III has also been tested in the CNA cyclotron (Seville) with 18 MeV proton beam to produce 4.439 MeV gamma rays. Data have been acquired with a graphite target in five different positions at 2.5 nA nominal beam intensity. Images with 4.439 MeV photons have been reconstructed, demonstrating the system capability to reconstruct images at energies relevant for hadron therapy. Moreover, the system has been able to distinguish 1mm displacements in the target position.