Back to Search Start Over

Highly selective potentiometric determination of Fe(III) ions using Tris-(1,2-diiminocyclohexylmethyl-5-Cl-2-hydroxyl benzaldehyde) based membrane electrode

Authors :
M Darvishpour
Mohammad Hossein Fekri
H. Dashti Khavidaki
Mahdieh Chegeni
Source :
JOURNAL OF ADVANCES IN CHEMISTRY. 11:3798-3803
Publication Year :
2015
Publisher :
CIRWOLRD, 2015.

Abstract

A novel ion-selective poly(vinyl chloride) membrane sensor for Fe(III) ions based on Tris-(1,2-diiminocyclohexylmethyl-5-Cl-2-hydroxyl benzaldehyde) are reported in this paper. The electrode exhibits a good potentiometric response for Fe(III), response time ≤ 20 s, over a wide concentration range 1.0 × 10-5 to 1.0 × 10-1 M with a slope 19.4 ± 0.5 mV/decade. The potentiometric response is independent on the pH of solution in the range of 1.5-5.0. The proposed electrode can be used for at least two months without any considerable divergence in potentials.It exhibits very good selectivity relative to a wide variety of alkali, alkaline earth, transition and heavy metal ions. The electrode assembly was also used as indicator electrode in the potentiometric titration of Fe(III) with EDTA.

Details

ISSN :
2321807X
Volume :
11
Database :
OpenAIRE
Journal :
JOURNAL OF ADVANCES IN CHEMISTRY
Accession number :
edsair.doi...........137fa98588093bbe43b5f04673828fbc
Full Text :
https://doi.org/10.24297/jac.v11i8.2187