Back to Search Start Over

A bat-shape piezoresistor electronic stethoscope based on MEMS technology

Authors :
Sicun Duan
Yuqi Lian
Weidong Wang
Yiming Shi
Lansheng Zhang
Renxin Wang
Xiaoyong Zhang
Guojun Zhang
Qingda Xu
Source :
Measurement. 147:106850
Publication Year :
2019
Publisher :
Elsevier BV, 2019.

Abstract

In this paper, a bat-shape electronic stethoscope based on MEMS technology is presented. The detector microstructure in bat-shape electronic stethoscope is designed by combining theoretical analysis and finite element method. Visual heart sound signal processing includes circuit and upper computer design. The frequency of the clinical heart sound signal is mostly in the range of 20–600 Hz. The sensitivity of the microstructure, its work-bandwidth and the SNR of the system are the key points of this design. And the final test results showed that the sensitivity of the sensor is −180.7 dB@500 Hz and SNR is 38 dB (17 dB higher than the 3 M electronic stethoscope). Based on theoretical analysis and test results, the bat-shape MEMS electronic stethoscope designed in this paper features small size, high sensitivity, high SNR and easy processing, and has broad prospects in the field of heart sound detection.

Details

ISSN :
02632241
Volume :
147
Database :
OpenAIRE
Journal :
Measurement
Accession number :
edsair.doi...........142a974611b35f5d76fbd9d3d1332033
Full Text :
https://doi.org/10.1016/j.measurement.2019.106850