Technical Complication of a Zirconia Multiple-Unit Fixed Dental Prosthesis Supported by Titanium Base Abutments—Case Report on a Bonding Failure and Treatment Alternative
- Source :
- The International Journal of Prosthodontics. 34:518-527
- Publication Year :
- 2021
- Publisher :
- Quintessence Publishing, 2021.
-
Abstract
- Monolithic zirconia implant-supported restorations connected to titanium bases or titanium inserts are increasing in popularity due to their application in a full digital workflow. These prefabricated abutments are connected to the all-ceramic superstructure by adhesive cementation. Although limited clinical data on the outcomes of this type of restoration are available, a few laboratory studies have shown possible debonding issues. This case report presents a bonding failure of a fixed dental prosthesis supported by titanium bases after short clinical use. A treatment alternative is also proposed using the available digital dental technology.