Development of a mobile radon calibration system at KRISS

Authors :
Byungjoo Kim
M.J. Han
Dong-Hye Heo
B. C. Kim
J.M. Lee
K.B. Lee
Y. G. Seon
Seung-Mi Hwang
Source :
Journal of Radioanalytical and Nuclear Chemistry. 330:571-576
Publication Year :
2021
Publisher :
Springer Science and Business Media LLC, 2021.

Abstract

A mobile radon calibration system was developed with a high-sensitivity radon detector, a movable calibration chamber, and a 226Ra solid source at the Korea Research Institute of Standards and Science (KRISS). The high-sensitivity radon detector was based on electrostatic collection mechanism. The volume of the calibration chamber was (1003.8 ± 6.0) L. The mobile radon calibration system was calibrated with 222Rn CRM from the radon primary system at KRISS. The sensitivities were established to be (231.8 ± 16.1) cpm/(kBq/m3) for 218Po and (254.9 ± 17.8) cpm/(kBq/m3) for 214Po, respectively (k = 2).