Back to Search Start Over

Effects of phase cancellation on broadband ultrasonic attenuation for in vitro human trabecular bone using a 2-D synthetic array

Authors :
Yi-Xian Qin
Frederick Serra-Hsu
Liangjun Lin
Wei Lin
Jiqi Cheng
Source :
2012 IEEE International Ultrasonics Symposium.
Publication Year :
2012
Publisher :
IEEE, 2012.

Abstract

The average BV/TV, TB.Th and Tb.Sp for the trabecular bone are 0.154±0.052, 0.203±0.025 mm and 0.812±0.151 mm, respectively. After normalization with sample thickness, the average normalized BUA (nBUA) is 24.8±9.5 dB/MHz/cm and 19.2±5.5 dB/MHz/cm for PS and PI respectively. PS nBUA is significantly different than PI BUA (paired t-test, p

Details

Database :
OpenAIRE
Journal :
2012 IEEE International Ultrasonics Symposium
Accession number :
edsair.doi...........1f14a7ce44c01938aaa7c8126e3f4eff
Full Text :
https://doi.org/10.1109/ultsym.2012.0278