Back to Search
Start Over
Eight-channel time measurement system with 2 ns resolution
- Source :
- Journal of Physics E: Scientific Instruments. 14:638-640
- Publication Year :
- 1981
- Publisher :
- IOP Publishing, 1981.
-
Abstract
- A microprocessor controlled time measurement system MNSC with eight separate stop channels started in groups of four channels is presented. The system uses integrated ECL counters with built-in gates as first stages counting a 500 MHz clock signal. The further counter stages are built up with STTL and LSTTL. Time intervals up to 20 ms can be measured with a resolution of 2 ns. The result of an experiment is shown where the arrival of a detonation front is measured with MNSC.
Details
- ISSN :
- 00223735
- Volume :
- 14
- Database :
- OpenAIRE
- Journal :
- Journal of Physics E: Scientific Instruments
- Accession number :
- edsair.doi...........280ba00f85a66f14b645b7b14c001d22
- Full Text :
- https://doi.org/10.1088/0022-3735/14/5/025