Back to Search Start Over

A Method for the Calibration of Synchrotron Beam Pickup Electrodes

Authors :
U. Vogel
A. I. Baron
Source :
Review of Scientific Instruments. 41:1832-1836
Publication Year :
1970
Publisher :
AIP Publishing, 1970.

Abstract

A method is described for the measurement of characteristics of pickup electrodes (PUE) for monitoring coordinates of an accelerator beam. A new representation is discussed for determining the coordinates of the center of charge for the entire PUE aperture. A few prototypes of PUE's to be installed in the AGS were tested and the results show that nonlinear deviations are at least 1 mm/25 mm. The method of measurement has a resolution of at least 0.1 mm/25 mm; it can therefore be used for (a) the design of high quality PUE's and (b) calibration of PUE's as installed in the accelerator in order to increase the usable aperture and the accuracy of beam coordinate monitoring.

Details

ISSN :
10897623 and 00346748
Volume :
41
Database :
OpenAIRE
Journal :
Review of Scientific Instruments
Accession number :
edsair.doi...........29a9e2b2e1e3a23d57cf221744075be4
Full Text :
https://doi.org/10.1063/1.1684421