Back to Search
Start Over
Experimental set-up for thermopower and resistivity measurements at 100-1300 K
- Source :
- Measurement Science and Technology. 12:264-272
- Publication Year :
- 2001
- Publisher :
- IOP Publishing, 2001.
-
Abstract
- In this paper we describe an experimental set-up for simultaneous measurements of thermopower and electrical resistivity at temperatures from 100 K to 1300 K. Optimal configuration of electrodes and original mechanical contacts of thermocouples, current leads and potential probes with the sample make it possible to measure a large variety of materials and result in greater flexibility with respect to the sample form and dimensions. Both bulk and thin film samples with resistances in the range from 0 Ω up to 200 kΩ (1 GΩ in case of the resistivity measurement) can be investigated. Precision and high reliability of the system have been proven during more than three years of use. The resistivity and thermopower of pure Pb, Cu and Ni, and a Cr-Si thin film composite are presented as test materials to demonstrate the possibilities and accuracy of this experimental set-up.
- Subjects :
- Materials science
Condensed matter physics
business.industry
Applied Mathematics
Electrical engineering
Measure (mathematics)
Reliability (semiconductor)
Thermocouple
Electrical resistivity and conductivity
Seebeck coefficient
Electrode
Thin film
business
Instrumentation
Engineering (miscellaneous)
Subjects
Details
- ISSN :
- 13616501 and 09570233
- Volume :
- 12
- Database :
- OpenAIRE
- Journal :
- Measurement Science and Technology
- Accession number :
- edsair.doi...........2b1cd1494bc6df49a7f1413f268daef4
- Full Text :
- https://doi.org/10.1088/0957-0233/12/3/304