The development of Campbell integration system applied in high particle flux measurement
- Source :
- Instruments and Experimental Techniques. 60:789-794
- Publication Year :
- 2017
- Publisher :
- Pleiades Publishing Ltd, 2017.
-
Abstract
- Campbell Mode is a method widely used in nuclear signal processing. In this article, we introduce the design of the new system based on the Campbell integration measurement applied in NFM (neutron flux measurement) systems for ITER plasma diagnose. In this article, we introduce a digital particle flux measurement system based on Campbell integration theory. A subsequent series of experiment are conducted to test the digital Campbell system performance, which proves that the Campbell integration system is competent for the plasma diagnose in ITER.
- Subjects :
- 010302 applied physics
Signal processing
Series (mathematics)
Computer science
business.industry
System of measurement
01 natural sciences
010305 fluids & plasmas
Development (topology)
Neutron flux
Condensed Matter::Superconductivity
Mathematics::Quantum Algebra
0103 physical sciences
Aerospace engineering
business
Particle flux
Instrumentation
Subjects
Details
- ISSN :
- 16083180 and 00204412
- Volume :
- 60
- Database :
- OpenAIRE
- Journal :
- Instruments and Experimental Techniques
- Accession number :