Back to Search
Start Over
Microfluidic paper-based device integrated with smartphone for point-of-use colorimetric monitoring of water quality index
- Source :
- Measurement. 164:108085
- Publication Year :
- 2020
- Publisher :
- Elsevier BV, 2020.
-
Abstract
- The measurements by lab-on-a-chip devices are very interesting regarding the reduction of reagents, time and cost. In this paper, we have described the construction and application of a new device based on colorimetric measurements using a smartphone camera and a microfluidic paper-based device (μPAD), both fixed in supports obtained through 3D printing, employing a polylactic acid filament. The μPAD was evaluated as a water quality index device by colorimetric monitoring of water hardness (Ca2+ and Mg2+), total phenols (catechol), and pH. Under optimized parameters, the proposed method presented a linear range between 20 and 560 mg L−1 for Ca2+ and Mg2+(hardness), 0.20 to 16 mg L−1 of catechol and pH from 4.7 to 12, and limits of detection of 0.083 mg L−1, 0.124 mg L−1 and 0.262, for water hardness, catechol, and pH, respectively. The proposed device was successfully applied for the quality monitoring of the distribution of water samples.
- Subjects :
- Detection limit
Materials science
Applied Mathematics
020208 electrical & electronic engineering
010401 analytical chemistry
Microfluidics
Analytical chemistry
02 engineering and technology
Paper based
Condensed Matter Physics
01 natural sciences
0104 chemical sciences
Linear range
Reagent
0202 electrical engineering, electronic engineering, information engineering
Point (geometry)
New device
Water quality
Electrical and Electronic Engineering
Instrumentation
Subjects
Details
- ISSN :
- 02632241
- Volume :
- 164
- Database :
- OpenAIRE
- Journal :
- Measurement
- Accession number :