Optimized illumination for high-throughput ptychography

Authors :
Jeffrey A. Klug
Yudong Yao
Zhonghou Cai
Christian Roehrig
Oliver Cossairt
Michael Wojcik
Stefan Vogt
Curt Preissner
Barry Lai
Junjing Deng
Yi Jiang
Youssef S. G. Nashed
Source :
X-Ray Nanoimaging: Instruments and Methods IV.
Publication Year :
2019
Publisher :
SPIE, 2019.

Abstract

As a scanning version of coherent diffraction imaging (CDI), X-ray ptychography has become a popular and very successful method for high-resolution quantitative imaging of extended specimens. The requirements of mostly coherent illumination and the scanning mechanism limit the throughput of ptychographic imaging. In this paper, we will introduce the methods we use at the Advanced Photon Source (APS) to achieve highthroughput ptychography by optimizing the parameters of the illumination beam. One work we have done is increasing the illumination flux by using a double-multilayer monochromator (DMM) optics with about 0.8% bandwidth. Compared with our double-crystal monochromator (DCM) optics with 0.01% bandwidth, this DMM optics provides around 20 times more flux. A multi-wavelength reconstruction method has been implemented to deal with the consequential degraded temporal coherence from such an illumination to ensure high-quality reconstruction. In the other work, we adopt a novel use of at-top focusing optics to generate a at-top beam with the diameter of about 1.5 μm on the focal plane. The better uniformity of the probe and the large beam size allow one to significantly increase the step size in ptychography scans and thereby the imaging efficiency.

Details

Database :
OpenAIRE
Journal :
X-Ray Nanoimaging: Instruments and Methods IV
Accession number :
edsair.doi...........3d450a4c3bf58300a3adf9184b75515e
Full Text :
https://doi.org/10.1117/12.2530527